OBJECTIVE: To establish a method for the determination of related substances in Cabazitaxel injection. METHODS: HPLC method was used. The determination was performed on Agilent Eclipse XDB-C18 column with mobile phase consisted of water-acetonitrile-ethanol (gradient elution) at the flow rate of 1.0 mL/min. The column temperature was 35 ℃, and the detection wavelength was set at 230 nm. The sample size was 20 μL. Established method was used to determine related substances in 3 batches of Cabazitaxel injection. RESULTS: The linear relationship of cabazitaxel were 0.039-11.60 μg/mL (r=0.999 8, n=7). The detection limit was 2×10-4 μg, and quantitation limit was 8×10-4 μg. RSD of precision and reproducibility tests were all lower than 10.0% (n=6). The amount of single impurity in 3 batches of samples ranged 0.07%-0.08%, and total amount of impurities were 0.26%-0.29%. CONCLUSIONS: Established method is simple, accurate and reliable, can be used for the determination of related substances in Cabazitaxel injection.