OBJECTIVE: To establish a method for simultaneous determination of residual methylparaben, ethylparaben, nipasol and benzalkonium chloride in marketed eye drops. METHODS: HPLC method was adopted. The determination was performed on Hypersil GOLD C18 column with mobile phase consisted of 0.005 mol/L ammonium acetate (10 mL triehtylamine in 1 L solution, pH adjusted to 5.0±0.5 with glacial acetic acid)-acetonitrile (45 ∶ 55,V/V) at the flow rate of 1.0 mL/min. The detection wavelength was 262 nm (methylparaben, ethylparaben, nipasol) and 214 nm (benzalkonium chloride), respectively. The column temperature was 30 ℃ and sample size was 20 μL. RESULTS: The linear range were 1.235 0-15.438 0 μg/mL for methylparaben(r=0.999 9), 1.317 0-16.383 6 μg/mL for ethylparaben (r=0.999 7), 1.207 2-15.089 4 μg/mL for nipasol (r=0.999 6) and 17.776-222.0 μg/mL for benzalkonium chloride (r=0.999 9), respectively. Limits of quantitation were 2.0, 2.0, 2.0, 1.11 μg, respectively; limits of determination were 0.375, 0.375, 0.375, 0.333 μg, respectively. RSDs of precision, stability and reproducibility tests were all lower than 2.0%. The average recoveries were 98.14%-102.48% (RSD=1.6%,n=9), 98.79%-102.42%(RSD=1.3%, n=9), 98.19%-102.49%(RSD=1.5%,n=9) and 98.76%-100.53% (RSD=0.6%,n=9), respectively. CONCLUSIONS: The method is accurate, reproducible, simple and suitable for the determination of residual methylparaben, ethylparaben, nipasol and benzalkonium chloride in marketed eye drops.