OBJECTIVE: To establish a method for simultaneous determination of BHT, free sulfur, antioxidant 1010, antioxidant 330, antioxidant 1076 and antioxidant 168 in pharmaceutical halogenated butyl rubber stopper. METHODS: HPLC method was adopted. The determination was performed on Agilent Zorbax SB-C18 column with mobile phase consisted of methanol-water (gradient elution) at the flow rate of 1.0 mL/min. The column temperature was 25 ℃, and detection wavelength was set at 220 nm. The sample size was 10 μL. RESULTS: The linear ranges of BHT, free sulfur, antioxidant 1010, antioxidant 330, antioxidant 1076 and antioxidant 168 were 2.592-64.80 μg/mL (r=0.999 9), 2.648-66.20 μg/mL (r=0.999 9), 2.062-51.55 μg/mL (r=0.999 9), 2.654-66.35 μg/mL (r=0.999 9), 2.128-53.20 μg/mL (r=0.999 9), 2.468-61.70 μg/mL (r=0.999 9), respectively. The limits of quantitation were 2.073 6, 2.118 4, 2.749 2, 1.061 6, 4.436 0, 3.090 4 mg/kg; the limits of detection were 0.628 4, 0.641 9, 0.833 1, 0.321 7, 1.344 2, 0.992 4 mg/kg, respectively. RSDs of precision, stability and reproducibility tests were all lower than 3%. The average recoveries were 96.57%-99.17%(RSD=0.92%, n=9), 77.47%-86.14%(RSD=2.94%, n=9), 96.11%-98.39%(RSD=0.63%, n=9), 96.22%-98.33%(RSD=0.67%, n=9), 95.59%-99.22%(RSD=1.11%, n=9) and 81.54%-88.31%(RSD=2.46%, n=9), respectively. CONCLUSIONS: The method is sensitive, rapid, convenient and reproducible, and can be used for simultaneous determination of 5 antioxidant and free sulfur in pharmaceutical halogenated butyl rubber stopper.