OBJECTIVE: To improve HPLC for content determination of ketoprofen in Ketoprofen enteric-coated capsules. METHODS: HPLC method was adopted. The determination was performed on Chiralpak IC column with mobile phase consisted of n-hexane (0.1% TFA)-isopropanol (90 ∶ 10,V/V) at a flow rate of 0.8 mL/min. The detection wavelength was set at 268 nm, and column temperature was 25 ℃. The sample size was 10 μL. RESULTS: The linear range of ketoprofen were 0.025-0.5 mg/mL(r=0.999 8). The limit of quantitation was 1.0 mg/L, and limit of detection was 0.2 mg/L. RSDs of precision, stability and reproducibility tests were lower than 2%; recoveries were 96.36%-100.32% (RSD=1.87%,n=6). CONCLUSIONS: The method is simple, accurate and rapid, and can be used for the content determination of ketoprofen in Ketoprofen enteric-coated capsules.